The LTD_1000 family of tags records a time series of temperature and
depth (absolute pressure) measurements to provide insight into detailed
animal behavior. Designed for fisheries management and research, these
tags offer a wide range of features and specifications to suit your
study's needs, from the extremely small and economical LTD_1100 series
to the extended life and deep diving LTD_1400 series.
- LTD_1100 available with 32K and 64K non-volatile memory.
- LTD_1110 available with 32K, 64K and 128K non-volatile memory.
- Standard depth measurement ranges include 50m, 500m and 1000m, with custom ranges available.
The
LTD_1100 series of temperature and depth recorders are surprisingly
small, extremely simple to use, and economical. Both external (LTD_1100)
and internal (LTD_1110) versions are available with a wide range
of depth measuring ratings and several memory configurations.
Several ingenious features are incorporated in this product, eliminating the user's requirement to configure the tag prior to deployment. These features also optimize tag performance, including memory usage and pressure resolution, as they adapt to operational conditions.
Time Extension
Recording
The configuration of tags normally requires anticipating the mission
length and the animal's behavior to set the appropriate sampling interval.
Time Extension Recording (TER) is an advanced memory management feature
that adjusts the sampling interval as recording proceeds so that,
regardless of the length of the recording session, the record is detailed
and the full session is recorded. TER continuously optimizes memory
usage over the entire deployment. Temperature and Pressure are sampled
at the same moment for ease of data analysis.
Here's how TER works: the logger begins recording at a base sampling interval of 1 sample every 14.063 seconds or 256 samples per hour. When its memory is full, the logger doubles its sampling interval and continues recording, overwriting every other sample in memory, and when its memory is filled at that sampling interval it doubles its sampling interval again. The process can repeat almost indefinitely. You can download a record of the full recording session at any time. As an option, a maximum interval can be specified at time of purchase to prevent unwanted dilution of the data's temporal resolution. The tag memory and sampling interval are reset using a magnet.
Depth Auto-Scaling
Depth Auto-Scaling automatically scales the pressure range to suit
the requirements of the application to provide finer resolution and
accuracy when needed. It initially records at a range of one-fourth
of the instrument maximum. If that range is exceeded, it switches
to half of the instrument maximum. If that is exceeded, it switches
to the maximum range. This process is non-reversible.
| LTD_1100 | LTD_1110 | |
| Dimensions | 8mm X 16mm X 27mm | 11mm X 32mm |
| Weight in air | 5 grams | 5 grams |
| Weight in Fresh Water | ~ 2 grams | ~ 2 grams |
| Memory | 32K, 64K | 32K, 64K, 128K |
| Memory Management | Time Extension Recording | Custom Programming |
| Survival Depth | 1000 meters | 1000 meters |
| Max Operation Temperature | 35° C | 35° C |
| Min Operation Temperature | -5° C | -5° C |
| Data Points (Temp and Pressure) | 16K, 32K | 16K, 32K, 64K |
| Optical Download | Yes, uses RDR1100 | Yes, uses RDR1100 |
| Data Download (Optical) | 19.2 kB/sec | 19.2 kB/sec |
| Standard Depth Range | 50, 500, 1000 m | 50, 500, 1000 m |
| Depth Auto-Scale | 1/4, 1/2, Full Scale | 1/4, 1/2, Full Scale |
| Data Resolution | 8 bits | 8 bits |
| Sensors | Temperature and Pressure | Temperature and Pressure |
| Pressure Accuracy | +/-1% of Current Scale | +/-1% of Current Scale |
| Pressure Resolution | Scale / 256 | Scale / 256 |
| Temperature Measurement | -5° to 35° C | -5° to 35° C |
| Temperature Accuracy | better than 0.3° C | better than 0.3° C |
| Temperature Resolution | 0.20° C (avg.) | 0.20° C (avg.) |
| User Setup | Not Required | Not Required |
| Time Extension Recording | Yes | Yes |
| Typical Life | >3 years | >2 years |
| Number of Records | 32K, 64K | 32K, 64K, 128K |
| Data Analysis Software | Included (TagTalk1100) | Included (TagTalk1100) |
The LTD_1100 series comes with data analysis software, TagTalk 1100, also available directly from our Downloads page.
LTD
1100 product sheet (PDF). ![]()
Avian
LTD products (PDF). ![]()